Type 15 Detector EFC Status:
Item Description
Type 15
Displays the Detector EFC type. Detector EFC Type 15
has 3 channels and is used for PFPD detector gas control.
Ready Yes or No.
Fault Yes or No.
Ch 1 (Air 1) Set,
Actual
Displays, in ml/min, the programmed and actual channel 2
flow rate.
Ch 2 (H2) Set, Actual Displays, in ml/min, the programmed and actual channel 2
flow rate.
Ch 3 (Air 2) Set,
Actual
Displays, in ml/min, the programmed and actual channel 2
flow rate.
Type 16 Detector EFC Status:
Item Description
Type 16
Displays the Detector EFC type. Detector EFC Type 16 has
2 channels and is used for TCD detector gas control.
Ready Yes or No.
Fault Yes or No.
Ch 1 (Make up) Set,
Actual
Displays, in ml/min, the programmed and actual channel 1
flow rate.
Ch 2 (Reference) Set,
Actual
Displays, in ml/min, the programmed and actual channel 2
flow rate if the channel 2 gas type is a Hydrogen reference.
Ch 3 (Make up) Set,
Actual
Displays, in ml/min, the programmed and actual channel 2
flow rate if the channel 2 gas type is a Hydrogen make-up.